Comparison of aerosol measurement systems during the 2016 airborne ARISTO campaign

2019 
AbstractSeveral different types of measurements of particle size and concentration were compared during the 2016 Airborne Research Instrumentation Testing Opportunity (ARISTO) campaign. The scanning mobility particle sizer (SMPS) measured number-size distributions for mobility diameters between ∼20–350 and ∼8–110 nm, depending on the mobility analyzer chosen. Also included were two stand-alone condensation particle counters (CPC) for determining size-integrated particle concentrations. A wing-mounted and a rack-mounted Ultra-High Sensitivity Aerosol Spectrometer (UHSAS) were used to measure size distributions between 60 and 1000 nm. Lastly, two different sampling inlets were used to investigate performance and observe any systematic biases. Most sampling occurred during cloud-free summer conditions in the western United States. Number concentrations from the two CPCs typically agreed within 12% once the flows in the ultrafine particle counter were corrected as a function of pressure. As expected, the size...