Improvement of Coagulation Laboratory Practice in Thailand: The First-Year Experience of the National External Quality Assessment Scheme for Blood Coagulation

2009 
Abstract Context.—In Thailand until 2005 there had been no external quality assessment scheme at the national level for blood coagulation tests. Only a few laboratories had an external quality assessment for these tests. In the year 2005, the Thailand National External Quality Assessment Scheme for Blood Coagulation was founded. Objectives.—To describe the establishment of the Thailand National External Quality Assessment Scheme for Blood Coagulation (including problems encountered and solutions), its progression and expansion, and the improvement of coagulation laboratory practice in Thailand during 2 trial surveys and 4 formal surveys conducted in the first 1;h1 years. Design.—Between 2005 and 2006, the external quality assessment samples for prothrombin time/international normalized ratio and activated partial thromboplastin time were distributed to the participants as well as the instructions and suggestions for the improvement of laboratory practice. From the data collected, the all-method coefficien...