High performance detectors for upgraded gamma ray diagnostics for JET DT campaigns
2016
In forthcoming deuterium-tritium (DT) experiments on JET a significant population of alpha-particles will be produced. For operating alpha-particle diagnostics at high DT neutron fluxes, specific improvements have to be made. Proposed new detectors for gamma-ray measurements will be based on CeBr3 and LaBr3:Ce scintillators. They are characterized by a good energy resolution, a relatively high detection efficiency for a few MeV gamma-rays and a fast response time. An overview of scintillator parameters is presented. A description of the properties of photodetectors is given to indicate optimal setups. Results of measurements, using gamma-ray sources with energies up to a few MeV, are discussed with relation to the DT campaign requirements.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
40
References
19
Citations
NaN
KQI