Evaluation of a commercially available carbohydrate deficient transferrin kit to detect beta-2-transferrin in cerebrospinal fluid using capillary electrophoresis.
2008
Abstract Objective The aim of this study was to evaluate the CEofix™ CDT Capillary Electrophoresis (CE) kit for the detection of β-2-tranferrin (β-2-Tf) in cerebrospinal fluid (CSF). Design and method Evaluation was performed according to CLSI EP5-A and EP12-A guidelines. Results The method resolved β-2-Tf from other Tf isoforms. The C50 for β-Tf was determined to be 0.46 mg/L. Neither hemoglobin nor bilirubin co-migrated with β-2-Tf. Conclusions The CDT kit can be used for detecting β-2-Tf in CSF by CE.
Keywords: