Air Embolism During TEVAR: An Additional Flush Port on the Delivery System Pusher Significantly Reduces the Amount of Air Released During Deployment of a Thoracic Stent-Graft in an Experimental Setting:

2018 
Purpose: To investigate the influence of (1) an additional side port for flushing the hollow pusher in Zenith thoracic stent-graft delivery systems and (2) additional carbon dioxide flushing on the amount of air released during stent-graft deployment. Methods: Twenty thoracic stent-grafts with an additional flush port to fill the hollow pusher were separated into 2 equal groups (C and D). Both groups were flushed with 20 mL of normal saline through the extra side port connected to the pusher and with 60 mL of saline through the regular flushing port. One group of grafts (group D) was additionally flushed with carbon dioxide through the regular flushing port prior to saline. All grafts were deployed into a curved plastic pipe attached to the bottom of a water-filled container. The released gas was recorded and measured using a calibrated setup. To evaluate the influence of the extra side port irrespective of the carbon dioxide flushing technique, group C was compared with a previously published reference g...