[Determination of manganese in urine by direct dilution-inductively coupled plasma mass spectrometry].
2021
Objective: To establish a direct dilution-inductively coupled plasma mass spectrometry method for the determination of manganese in urine. Methods: Using 1% nitric acid solution as diluent, the urine dilution factor and internal standard elements were determined by single factor rotation experiment. The linear range, correlation coefficient, precision, accuracy and detection limit of the direct dilution-inductively coupled plasma mass spectrometry for the determination of manganese in urine were evaluated. Results: The linear range of this method was 0.0-20 μg/L, the correlation coefficient was 0.999 9, the detection limit was 0.02 μg/L, the recoveries were 84.65%-103.40%, the relative standard deviations were 0.26%-8.17%. Conclusion: This method has the advantages of simple operation, high sensitivity and low detection limit. It can be used for the determination of urine manganese at the same time with other elements. It is suitable for the determination of urine manganese in workers and ordinary people.