A rapid method for measuring intraocular lens power in vitro with a focimeter

2015 
In this paper we describe a new method for measuring the intraocular lens (IOL) power using a focimeter, a negative ophthalmic lens and a saline solution (0.9% NaCl). To test this we measured the power of 58 different IOLs and we compared them with the power stated by the manufacturer. Despite the limitations, the results show a good correlation.