Performance of Shear Wave Elastography in Delineating the Radiofrequency Ablation Boundary: An in Vivo experiment

2019 
Abstract This study was aimed at exploring the cutoff value of Young's modulus of ablated tissue and the optimal scale at which shear wave elastography (SWE) can delineate the ablation boundary. The livers of 30 rabbits were radiofrequency (RF) ablated, and ultrasonic imaging, including SWE and contrast-enhanced ultrasound (CEUS), was performed. The ablation boundary in the SWE image was located using CEUS, and the SWE parameters of the boundary were measured to calculate the cutoff value of Young's modulus. The cutoff value of the ablated tissue was 48–50 kPa 2 h to 28 d post-ablation. The regions of increased stiffness in SWE images at a scale of 0–50 kPa overlapped well with the non-enhanced regions of CEUS images in 88% of specimens. Therefore, elasticity values differed significantly between ablated and non-ablated tissues, and the cutoff value for Young's modulus differentiated these tissues. SWE delineated the ablation boundary well at the optimal SWE scale with respect to the cutoff value.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    29
    References
    1
    Citations
    NaN
    KQI
    []