Optimization of stent implantation using a high pressure inflation protocol

2016 
Background High-pressure inflation is the universal standard for stent deployment but a specific protocol for its use is lacking. We developed a standardized “pressure optimization protocol” (POP) using time to inflation pressure stability as an endpoint for determining the required duration of stent inflation. Objectives: The primary study purpose was to determine the stent inflation time (IT) in a large patient cohort using the standardized inflation protocol, to correlate various patient and lesion characteristics with IT, and ascertain in an in vitro study the time for pressure accommodation within an inflation system. Methods Six hundred fifteen stent implants in 435 patients were studied. Multivariate analysis was performed to determine predictors of longer ITs. In an in vitro study, various stents and balloons were inflated in air to determine the pressure accommodation time of the inflation system. Results The mean stent IT was 104 ± 41 sec (range 30–380 sec). Stent length was the only predictor of prolonged stent inflation. The “accommodation time” in vitro of the stent inflation system itself was 33 ± 24 sec. The protocol was safe requiring premature inflation termination in <3% of stent implants. No serious adverse events occurred. Conclusions Achieving stable inflation pressure requires on average over 100 sec and may require several minutes in individual cases. Stent length increases IT. These results suggest that the widespread practice of rapid inflation/deflation may not be sufficient to fully expand the stent and that the use of a pressure stability protocol will allow for safe, predictable, and more complete stent deployment. © 2015 Wiley Periodicals, Inc.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    19
    References
    9
    Citations
    NaN
    KQI
    []