On-Line Electrodialytic Salt Removal in Electrospray Ionization Mass Spectrometry of Proteins
2011
Salts and buffers, commonly used in isolation and stabilization of biological analytes, have a deleterious effect on electrospray ionization mass spectrometry (ESI-MS). Excessive concentrations of salts lead to ion suppression and adduct formation, which mask or complicate ion signals. In this work, we describe a salt remover (SR), configured as a three-compartment flow-through system, where the central compartment is separated from the outer compartments by a cation-exchange membrane (CEM) and an anion-exchange membrane (AEM). One platinum electrode is placed in each of the outer compartments, where water or electrolyte is flowing. The CEM electrode is held at a negative potential relative to the AEM side; cations/anions migrate by electrophoresis to the CEM/AEM receiver liquids, respectively. Proteins have poorer electrophoretic mobility relative to the buffer components, permitting removal of the salt. The salt-free proteins proceed to the ESI source. The capillary scale SR (internal volume 2.5 μL) des...
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
42
References
33
Citations
NaN
KQI