Purpose: To assess correlation for anterior chamber flare grading between clinicians with different levels of experience and with semi-automated flare reading in a cohort of patients with heterogeneous uveitic entities. Methods: Fifty-nine observations from 36 patients were recorded and analyzed for statistical association. In each patient, flare was assessed objectively using the Kowa FM-700 laser flare photometer, and subjective masked grading by two clinicians was performed. Results: The study demonstrated disparity in flare readings between clinical graders with one step disagreement in clinical grading in 26 (44.06%) eyes (p50.001) and concordance between the flare readings by experienced grader and flare photometry. After review of semi-automated flare readings, management was changed in 11% of the patients. Conclusion: Laser flare photometry can be a valuable tool to remove the observer bias in grading flare for selected cohort of uveitis patients. It can be further applied to titrate therapy in intraocular inflammation.