Association between metal hypersensitivity and implant failure in patients who underwent titanium cranioplasty

2018 
OBJECTIVEDigitally designed titanium plates are commonly used for the reconstruction of craniofacial defects, although implant exposure (referred to as failure) is one of the major complications. Metal hypersensitivities have been suggested as possible causes of implant failure of orthopedic, intravascular, gynecological, and dental devices, yet there has been no consensus on the requirement for allergy screening before cranioplasty.METHODSIn this study, the authors prospectively investigated the prevalence of metal hypersensitivity in patients for whom cranioplasty is planned and assess its relationship with titanium implant failure (exposure).RESULTSBased on records from 207 included patients, 39.61% of patients showed hypersensitivity to at least one kind of metal. Approximately one-quarter (25.12%) of patients had multiple metal allergies. Co, Cd, and Zn were the 3 most frequently identified metal hypersensitivities. No allergy to titanium was detected in this study. The overall incidence of craniopla...