High-Frequency Irreversible Electroporation: Safety and Efficacy of Next-Generation Irreversible Electroporation Adjacent to Critical Hepatic Structures:

2017 
Irreversible electroporation (IRE) is a nonthermal ablation modality employed to induce in situ tissue-cell death. This study sought to evaluate the efficacy of a novel high-frequency IRE (H-FIRE) system to perform hepatic ablations across, or adjacent to, critical vascular and biliary structures. Using ultrasound guidance H-FIRE electrodes were placed across, or adjacent to, portal pedicels, hepatic veins, or the gall bladder in a porcine model. H-FIRE pulses were delivered (2250 V, 2-5-2 pulse configuration) in the absence of cardiac synchronization or intraoperative paralytics. Six hours after H-FIRE the liver was resected and analyzed. Nine ablations were performed in 3 separate experimental groups (major vessels straddled by electrodes, electrodes placed adjacent to major vessels, electrodes placed adjacent to gall bladder). Average ablation time was 290 ± 63 seconds. No electrocardiogram abnormalities or changes in vital signs were observed during H-FIRE. At necropsy, no vascular damage, coagulated-...