Comparison of mid-infrared and Raman spectroscopy in the quantitative analysis of serum

2005 
Mid-infrared or Raman spectroscopy together with multi- variate data analysis provides a novel approach to clinical laboratory analysis, offering benefits due to its reagent-free nature, the speed of the analysis and the possibility of obtaining a variety of information from one single measurement. We compared mid-infrared and Raman spectra of the sera obtained from 247 blood donors. Partial least squares analysis of the vibrational spectra allowed for the quantifica- tion of total protein, cholesterol, high and low density lipoproteins, triglycerides, glucose, urea and uric acid. Glucose (mean concentra- tion: 154 mg/dl) is frequently used as a benchmark for spectroscopic analysis and we achieved a root mean square error of prediction of 14.7 and 17.1 mg/dl for mid-infrared and Raman spectroscopy, re- spectively. Using the same sample set, comparable sample through- put, and identical mathematical quantification procedures Raman and mid-infrared spectroscopy of serum deliver similar accuracies for the quantification of the analytes under investigation. In our experiments vibrational spectroscopy-based quantification appears to be limited to accuracies in the 0.1 mmol/l range. © 2005 Society of Photo-Optical Instrumen-
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    26
    References
    96
    Citations
    NaN
    KQI
    []