Automated On-Line Isotope Dilution Analysis with ICP-MS Using Sandwich Flow Injection.

1998 
An automated flow injection (FI) manifold is described to perform the addition of isotopic spikes to aqueous samples on-line with ICP-MS for isotope dilution (ID) analysis. The manifold uses the sandwich technique (with the nested loop approach) to perform an injection of the isotopic spike solution within a sample (or standard) plug, the resulting sample-spike-sample sequence being pushed toward the nebulizer by a 1% HNO 3 carrier. A standard, which must contain one element not present in the spike solution to allow the determination of the dispersion coefficient, must also be used to allow a reverse isotope dilution analysis, as well as corrections for mass discrimination and/or spectroscopic interferences. Indeed, because the signals from the individual isotopes are monitored continuously, only one isotope free of spectroscopic interference is required for elements whose isotopic distribution does not vary in nature (two isotopes are still needed for the other elements), as a correction for the interference can be made by comparison with the signals from the standard. Furthermore, this automated approach makes ID-ICP-MS a faster method and does not require any preliminary analysis of the sample because the concentration profile resulting from FI allows the selection of the best isotopic ratio. It was successfully applied to the determination of Mo in saline water.