Atrial Tachyarrhythmia Recurrence Temporal Patterns in Bradycardia Patients Implanted with Antitachycardia Pacemakers
2004
Introduction: New-generation pacemakers allow continuous atrial tachyarrhythmia (AT) monitoring that provides accurate information about AT type, frequency, burden, and temporary evolution.
Methods and Results: We performed a prospective multicenter study to describe AT temporal patterns in patients with sinus bradycardia and AT. Two hundred forty patients (123 men; age 71 ± 8 years) were implanted with a DDDRP pacemaker (model AT500, Medtronic Inc.). All patients were followed for 13 months. The first-month stabilization period of all patients was discarded from analysis. Seventy percent of patients had AT recurrences. Mean time to first AT recurrence (48.2 days, 95% confidence interval [CI] 37.0–59.5 days) was significantly longer than the time between first and second AT episode (10.3 days, 95% CI 6.7–13.9 days, P 25 AT episodes and fitted by power law and exponential functions. The best fit was obtained by power law function in 60% of patients, by exponential function in 10%, and the two models gave comparable results in 30% of patients.
Conclusion: In our population of patients with a history of sinus bradycardia and AT who were implanted with a new device equipped with atrial pacing therapies, 30% did not experience AT recurrences in the 12-month study period. Analysis of interevent time showed that in 60% of patients AT recurrences do not follow a uniform or random distribution. These findings bring into question the use of cross-over design and time to first AT recurrence as a clinical outcome in trials for AT therapy in this patient population. (J Cardiovasc Electrophysiol, Vol. 15, pp. 44-51, January 2004)
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
28
References
24
Citations
NaN
KQI