Reliability of implantable cardioverter defibrillator home monitoring in forecasting the need for regular office visits, and patient perspective. Japanese HOME-ICD study.

2013 
Abstract Remote monitoring (RM) technology has emerged as a potentially efficient method to manage patients with implantable cardioverter defibrillators (ICDs) or cardiac resynchronization therapy defibrillators (CRT-Ds). This study evaluated the reliability of daily RM in forecasting the need for regular in-hospital follow-ups (RFUs). Two hundred and fifteen patients implanted with Biotronik Lumax devices (142 ICDs, 73 CRT-Ds) were enrolled. RFU was performed at 3, 6, 9, and 12 months after implantation. Immediately before an RFU, the physician forecasted the need for RFU based on RM data (pre-RFU assessment). A completed RFU session was classified as necessary if an action was undertaken potentially influencing patient safety, device therapy, or medication therapy (post-RFU assessment). Overall, 663 pairs of pre- and post-RFU assessments were compared. The number of pre-RFU assessments failing to predict the need for RFU was 38 (5.7%), fulfilling the study hypothesis of 5.0±4.0% (P