Absolute flatness measurements by Fizeau interferometer and 3D scanning deflectometric profiler
2020
The National Metrology Institute of Technology provides a flatness calibration service for large flat substrates using a Fizeau interferometer with 10 nm (k = 2) uncertainty over a measurement range of 300 mm. In the Fizeau interferometer, the surface profile of the reference flat is a measurement error factor because the measurand is the gap distance between the reference flat and the specimen. We have reduced uncertainty by constructing a surface profile map of the reference flat based on the three-flat test; however, it is difficult to manufacture a highly accurate reference flat and create a correction surface profile map in order to measure even larger flat substrates. Thus, we developed a three-dimensional scanning deflectometric profiler (3-D SDP) that does not require a reference flat and can directly measure a surface profile. Measuring devices based on deflectometry have been developed by many laboratories for highly accurate straightness profile measurement, but measurement by such systems is limited to a line (two-dimensional) profile. To solve this problem, we developed a novel method of measuring the surface topography by calculating the topography from radial lines obtained by rotating the specimen. In this study, we compared measurements between the Fizeau interferometer and 3-D SDP. We also report the results of measuring an optical flat with a diameter of 300 mm.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
0
References
0
Citations
NaN
KQI