Single- and dual-source chest CT protocols: Levels of radiation dose in routine clinical practice.

2012 
Abstract Purpose To establish the radiation dose level for single- and dual-source thoracic CT scans in daily practice. Materials and methods The dose levels delivered during 634 consecutive examinations over a period of 2 months were recorded. The CT scans were performed using: (a) a standard protocol (single source, single energy [group 1]: n  = 266; dual source, single energy [group 2]: n  = 276; (b) with prospective ECG synchronisation [group 3]: n  = 13; or (c) with dual energy [group 4]: n  = 79. All the acquisitions included kilovoltage selection depending on the weight and automatic milliamperage modulation. Results The mean DLP of the standard protocols was 97.12 mGy cm (group 2; BMI = 23.1 kg/m 2 ) and 211.1 mGy cm (group 1; BMI = 27.3 kg/m 2 ), the choice of protocol depending on the diameter of the thorax relative to the diameter of the field of the second source, and therefore on the patient's morphotype. When imaging included examination of the proximal and middle coronary arteries (group 3), the mean DLP was 105.5 mGy cm. Morphological and functional imaging (group 4) was obtained with a mean DLP of 404.3 mGy cm. Conclusion Depending on the objective of the protocol, the mean DLP varied from 97.12 to 404.3 mGy cm.