Improved Outcomes in Auditory Brainstem Implantation with the Use of Near-Field Electrical Compound Action Potentials
2014
ObjectiveTo compare the outcomes (auditory threshold and open-set speech perception at 48-month follow-up) of a new near-field monitoring procedure, electrical compound action potential, on positioning the auditory brainstem implant electrode array on the surface of the cochlear nuclei versus the traditional far-field electrical auditory brainstem response.Study DesignRetrospective study.SettingTertiary referral center.Subjects and MethodsAmong the 202 patients with auditory brainstem implants fitted and monitored with electrical auditory brainstem response during implant fitting, 9 also underwent electrical compound action potential recording. These subjects were matched retrospectively with a control group of 9 patients in whom only the electrical auditory brainstem response was recorded. Electrical compound action potentials were obtained using a cotton-wick recording electrode located near the surface of the cochlear nuclei and on several cranial nerves.ResultsSignificantly lower potential thresholds ...