Study of functional viability of SU-8-based microneedles for neural applications

2009 
This paper presents the design, fabrication, packaging and first test results of SU-8-based microneedles for neural applications. By the use of photolithography, sputtering and bonding techniques, polymer needles with integrated microchannels and electrodes have been successfully fabricated. The use of photolithography for the patterning of the fluidic channel integrated in the needle allows the design of multiple outlet ports at the needle tip, minimizing the possibility of being blocked by the tissue. Furthermore, the flexibility of the polymer reduces the risk of fracture and tissue damage once the needle is inserted, while it is still rigid enough to allow a perfect insertion into the neural tissue. Fluidic and electric characterization of the microneedles has shown their viability for drug delivery and monitoring in neural applications. First drug delivery tests in ex vivo tissue demonstrated the functional viability of the needle to deliver drugs to precise points. Furthermore, in vivo experiments have demonstrated lower associated damages during insertion than those by stereotaxic standard needles.