CALIBRATION OF CRYOGENIC THERMOMETERS FOR THE LHC

2008 
6000 cryogenic temperature sensors of resistive type covering the range from room temperature down to 1.6 K are installed on the LHC machine. In order to meet the stringent requirements on temperature control of the superconducting magnets, each single sensor needs to be calibrated individually. In the framework of a special contribution, IPN (Institut de Physique Nucleaire) in Orsay, France built and operated a calibration facility with a throughput of 80 thermometers per week.After reception from the manufacturer, the thermometer is first assembled onto a support specific to the measurement environment, and then thermally cycled ten times and calibrated at least once from 1.6 to 300 K. The procedure for each of these interventions includes various measurements and the acquired data is recorded in an ORACLE®-database. Furthermore random calibrations on some samples are executed at CERN to crosscheck the coherence between the approximation data obtained by both IPN and CERN. In the range of 1.5 K to 30 K,...