An improved reference measurement procedure for triglycerides and total glycerides in human serum by isotope dilution gas chromatography-mass spectrometry.

2014 
Abstract Background Triglycerides are widely tested in clinical laboratories using enzymatic methods for lipid profiling. As enzymatic methods can be affected by interferences from biological samples, this together with the non-specific nature of triglycerides measurement makes it necessary to verify the accuracy of the test results with a reference measurement procedure. Several such measurement procedures had been published. These procedures generally involved lengthy and laborious sample preparation steps. In this paper, an improved reference measurement procedure for triglycerides and total glycerides was reported which simplifies the sample preparation steps and greatly shortens the time taken. Methods The procedure was based on isotope dilution gas chromatography–mass spectrometry (ID GC–MS) with tripalmitin as the calibration standard. Serum samples were first spiked with isotope-labeled tripalmitin. For the measurement of triglycerides, the serum samples were subjected to lipid extraction followed by separation of triglycerides from diglycerides and monoglycerides. Triglycerides were then hydrolyzed to glycerol, derivatized and injected into the GC–MS for quantification. For the measurement of total glycerides, the serum samples were hydrolyzed directly and derivatized before injection into the GC–MS for quantification. Results All measurement results showed good precision with CV