Feasibility study on hybrid medical imaging device based on Compton imaging and magnetic resonance imaging

2009 
Abstract In this paper, we propose a combined Compton camera/magnetic resonance imaging (MRI) scanner. For this, the table-top Compton camera currently under development in our laboratory is suitable, considering that it is not very large (i.e., a table-top size) and that it uses semiconductor detectors (for both the scatterer and absorber detectors), which in principle are not very sensitive to a magnetic field. The Compton camera takes three-dimensional images from a fixed position and, therefore, does not require a large ring-type structure, making it possible to fit it into an existing MRI system, without requiring major modifications to the system. In the present study, the potential of combining the table-top Compton camera and an MRI scanner for real simultaneous imaging was demonstrated by fusing a Compton camera image of an instance of multi-tracing, generated by using Geant4 Monte Carlo simulations, with an MR image.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    17
    References
    4
    Citations
    NaN
    KQI
    []