Thirteen-year results of cementless total hip arthroplasty with zirconia heads on highly cross-linked polyethylene.

2020 
OBJECTIVES Zirconia ceramic was introduced for femoral heads because of its mechanical strength and fracture toughness. Highly cross-linked polyethylene has been shown to achieve wear reduction compared to conventional polyethylene. The purpose of this study was to investigate the long-term results of cementless total hip arthroplasty with zirconia femoral heads on highly cross-linked polyethylene. METHODS Radiographic measurements of two-dimensional femoral head penetration into the polyethylene were performed with a computerized method (PolyWare) in 93 hips at a mean follow-up of 13 years. The steady-state wear rates were measured based on the radiographs from the first year after surgery to the final follow-up. RESULTS The mean Merle d'Aubigne and Postel score had improved significantly from 10.9 points preoperatively to 15.3 postoperatively (p < 0.001). Osteolysis was not found around the implant, and no zirconia heads had fractures. The total head penetration rate was 0.04 mm/year, and the steady-state wear rate was 0.02 mm/year. Kaplan-Meier survivorship with the endpoint of revision was 97% (95% confidence interval [CI], 92% to 99%) at 10 years. CONCLUSIONS Zirconia on highly cross-linked polyethylene demonstrated extremely low wear properties at long-term follow-up.