Validation and clinical evaluation of a UHPLC method with fluorescence detector for plasma quantification of doxorubicin and doxorubicinol in haematological patients

2014 
Abstract A rapid and simple UHPLC–fluorescence detection method for the quantification of doxorubicin and its main metabolite, doxorubicinol, in human plasma has been developed. The method was also validated for its application in therapeutic drug monitoring, a clinical approach used in the optimization of oncologic treatments. Following a single protein precipitation step, chromatographic separation was achieved using a C 18 column (50 mm × 2.10 mm, particle size 1.7 μm) at 50 °C with a mobile phase consisting of water (containing 0.4% triethylamine and 0.4% orthophosphoric acid)/acetonitrile (77:23, v/v). Flow rate was 0.50 mL/min and fluorescence detection with an excitation wavelength of 470 nm and an emission wavelength of 548 nm was used. The method met the specifications of linearity, selectivity, sensitivity, accuracy, precision and stability of the FDA and EMA guidelines for the validation of bioanalytical methods. Linearity for the drug (8–3000 ng/mL) and the metabolite (3–150 ng/mL) was observed ( R 2  > 0.992) and the maximum intra-day and inter-day precision coefficients of variation were less than 14% for both. The lower limits of quantification were 8 and 3 ng/mL for doxorubicin and doxorubicinol, respectively. The method was successfully applied to the quantify plasma concentrations of doxorubicin and doxorubicinol in 33 patients diagnosed with haematological malignancies in which broad ranges for drug (8.3–2766.0 ng/mL) and metabolite (4.8–104.9 ng/mL) levels were measured adequately.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    22
    References
    14
    Citations
    NaN
    KQI
    []