A technique to measure the absorbed dose in human tooth enamel using EPR method

2008 
The EPR spectrum of irradiated tooth enamel contains a multitude of signals that are divided into two categories of radiation-induced and radiation insensitive (native) signals. At lower doses the broad native signal obscures the radiation-induced signal. In this work attempt has been made to find a method to measure the radiation-induced signal other than peak-to-peak signal amplitude measurement. For this reason software was programmed to extract the data from EPR system. The average amplitude of the radiation-induced EPR signal which is defined between the known g-values can also be calculated using the software. The result of this calculations were considered as the EPR response for the tooth enamel samples irradiated from 100 to 500 mGy and was drawn as the calibration curve. The resulted data as compared to the peak-to-peak amplitude measurement method seems to be more reproducible and shows a better variation against the dose values.