Simultaneous Determination of Losartan and Hydrochlorothiazide in Human Plasma by LC/MS/MS with Electrospray Ionization and Its Application to Pharmacokinetics

2009 
A method based on a simple liquid-liquid extraction (LLE) followed by high-performance liquid chromatography with negative ion electrospray ionization tandem mass spectrometry (HPLC-ESI-MS/MS) detection was developed for the simultaneous determination of losartan (LOS) and hydrochlorothiazide (HCTZ) in human plasma, using valsartan (VAL) and chlorthalidone (CHTD) as an internal standard, respectively. The acquisition was performed in multiple reactions monitoring (MRM) and the limit of quantification was 4 ng/mL for both LOS and HCTZ. The method was linear in the studied range (4–800 ng/mL for LOS and 4–500 ng/mL for HCTZ). The intra-assay precisions ranged from 2.6–11.9% for LOS and 1.4–8.2% for HCTZ, while the inter-assay precisions ranged from 1.0–8.0% for LOS and 2.5–7.7% for HCTZ. The intra-assay accuracies ranged from 91.3 to 107.6% for LOS and 91.5 to 105.8% for HCTZ, while the inter-assay accuracies ranged from 99.9 to 106.4% for LOS and 97.4 to 101.4% for HCTZ. The analytical method was applied t...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    24