Performance of the BacT/Alert Virtuo Microbial Detection System for the culture of sterile body fluids: a prospective multi-centered study

2017 
Abstract Objectives Continuous monitoring blood culture systems are commonly used for sterile body fluid cultures. In this multicentre study, we evaluated the performance of the new-generation BacT/Alert Virtuo system compared to the BacT/Alert 3D and conventional culture for the recovery of microorganisms from sterile body fluids. Methods Peritoneal, cerebrospinal, pericardial, pleural and synovial fluids from adult patients submitted for culture were collected from three different centres. Specimens were inoculated into two bottles of the same bottle type (SA, SN, FA Plus or FN Plus) in equal volumes for simultaneous incubation in the Virtuo and 3D instruments. Each specimen was also Gram stained and seeded to solid media. Results A total of 811 specimens were inoculated to 1257 bottle pairs. The Virtuo and 3D showed equivalent recovery of clinically significant microorganisms (127/155, 81.9%, vs. 126/155, 81.3%, respectively). Solid media cultures recovered fewer pathogens than either continuous monitoring system (95/155, 61.3%, p  Enterobacteriaceae and enterococci. The Virtuo was significantly faster than the 3D in median time to detection of isolates from the same specimen (12.5 (range, 2.8–101.5) hours vs. 15.5 (range, 4.3–78.5) hours, p  Conclusions The BacT/Alert Virtuo system was equivalent to the 3D system in organism recovery from sterile body fluid culture but showed faster detection of growth as a result of design enhancements.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    21
    References
    7
    Citations
    NaN
    KQI
    []