The Root-Cause Analysis on Failed Patient-Specific Measurements of Pencil-Beam-Scanning Protons.

2021 
Purpose/Objective(s) The purpose of this study is to present the root-cause analysis on failed patient-specific quality assurance (QA) measurements of pencil-beam-scanning (PBS) protons; referred to as PBS-QA measurement. Materials/Methods A criterion to fail a PBS-QA measurement is having a Results A total 2488 PBS-QA measurements were catalogued. The percentage of measurements for the sites of Head/Neck, Breast, Prostate, and Other are 53.3%, 22.7%, 10.5%, and 13.5% respectively. The percentage of measurements with a passing rate of 100%-to-95%, 94%-to-89%, and 3% per-mm dose gradient along depth of beam path was observed on the failed measurements. To study the effect of depth dose gradient, a 3D Gamma-index analysis toolkit was used. 43.1% of measurements improved using this 3D toolkit. A feature among measurements that remained sub-optimal was a sharp > 3% per-mm lateral dose gradient that may not be well handled by using the detector size of 5.0mm in-diameter. An analysis of the sampling of finite-size detectors using 1D error function showed a large dose deviation at locations of low doses in a valley between two plateaus. User error, large depth dose gradient, and the effect of detector-size are identified. Conclusion With an understanding and mitigation of these root causes, the goals of patient-specific QA, specifically detecting actual deviation of beam delivery or identifying limitations of dose calculation algorithm in used treatment planning system, can be directly related to the failure of PBS-QA measurements.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    0
    Citations
    NaN
    KQI
    []