A novel technique for ligation of the cephalic vein reduces hemorrhaging during a two-in-one insertion of dual cardiac device leads
2018
Abstract The cutdown technique for the cephalic vein is a common access route for transvenous cardiac device leads (TVLs), and sometimes one cephalic vein can accomodate two TVLs. We examined a novel ligation technique to balance the hemostasis and lead maneuverability for this two-in-one insertion. A total of 22 patients scheduled for cardiac device implantations with two or more leads were enrolled. The ipsilateral cephalic vein was identified for inserting the TVLs with a cutdown. If two TVLs could be introduced into one cephalic vein, hemostasis was established by ligating the venous wall between the TVLs. We measured the amount of hemorrhaging per minute and the operators assessed the lead maneuverability before and after the ligation. We successfully implanted cardiac devices in 15 patients (68%) with this novel method, whereas only one TVL could be introduced via the cephalic vein in 7 patients. As for the successful patients, hemorrhaging from the gap was significantly reduced (5.6 ± 7.3 to 0.41 ± 0.36g/min, p = 0.016) after the novel ligation. The lead maneuverability was well maintained so there was no difficulty placing the leads into the cardiac chambers in all cases. No major complications were observed. In the present study, the novel ligation method provided significant hemostasis as well as a preserved maneuverability. It could be an optional choice for insertion of multiple TVLs.