Final report on key comparison CCQM-K96: Determination of amount content of dichromate

2013 
The key comparison CCQM-K96 was organized jointly by the inorganic and electrochemistry working groups of CCQM to test the abilities of the metrology institutes to measure the amount content of dichromate. Slovak Institute of Metrology with help of KRISS acted as the coordinating laboratories. Eight NMIs took part in the comparison. All participants used high accuracy constant current coulometry. Good agreement of the results was observed. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).