Image Quality of Iodine Maps for Pulmonary Embolism: A Comparison of Subtraction CT and Dual-Energy CT

2019 
OBJECTIVE. The objective of this study was to compare the image quality of iodine maps derived from subtraction CT and from dual-energy CT (DECT) in patients with suspected pulmonary embolism (PE). SUBJECTS AND METHODS. In this prospective study conducted between July 2016 and April 2017, consecutive patients with suspected PE underwent unenhanced CT at 100 kV and dual-energy pulmonary CT angiography at 100 and 140 kV on a dual-source scanner. The scanner was set to generate subtraction and DECT iodine maps at similar radiation doses. In 55 patients (30 women, 25 men; mean age ± SD, 63.4 ± 11.9 years old), various subjective image quality criteria including diagnostic acceptability were rated on a 5-point scale by four radiologists and a radiology resident. In 29 patients (17 women, 12 men; mean age, 62.4 ± 11.7 years old) with confirmed perfusion defects, the signal-difference-to-noise ratio (SDNR) between perfusion defects and adjacent normally perfused parenchyma was measured in corresponding ROIs on s...