Evaluation of Commercial Molecular Diagnostic Methods for the Detection and Determination of Macrolide Resistance in Mycoplasma pneumoniae.

2020 
We evaluated six commercial molecular tests targeting M. pneumoniae: the BioFire FilmArray Respiratory Panel (RP), the Meridian Alethia Mycoplasma Direct, the GenMark ePlex Respiratory Pathogen Panel (RPP), the Luminex NxTAG RPP, the ELITech ELITe InGenius Mycoplasma MGB Research Use Only Polymerase Chain Reaction (PCR), and the SpeeDx Resistance Plus MP. Laboratory-developed PCR assays at the University of Alabama at Birmingham and the Centers for Disease Control and Prevention were used as reference standards. Among 428 specimens, 212 were designated confirmed-positives for M. pneumoniae. The highest clinical sensitivities were found with the InGenius (99.5%) and the FilmArray RP (98.1%). The Resistance Plus MP identified 93.3% of the confirmed-positive specimens, whereas 83.6%, 64.6%, and 55.7% were identified by the ePlex RPP, NxTAG RPP, and Mycoplasma Direct assays, respectively. There was no significant difference between the sensitivity of the reference methods and that of the FilmArray RP and InGenius assays, but the remaining four assays detected significantly fewer positive specimens (p