Quantitative assessment of diastolic filling function: A new Doppler echocardiographic method and in vitro validation

1996 
Abstract The effects of changes in preload and chamber compliance on the transmitral flow pattern was investigated with a fluid dynamics model. A decrease in compliance led to a steepening of the deceleration slope and a shortening of the pressure half-time in both restricted and nonrestricted valves. An increase in the preload led to a longer pressure half-time but did not affect the slope in the restricted valve. However, increased preload led to a shorter pressure half-time and a steeper slope in the nonrestricted valve. A mathematically derived method to measure the net atrioventricular compliance (sol-[annular area] va[ϱ vt [flow deceleration rate]]), where v a and v t are flow velocities at the mitral anulus and the mitral tip levels) was then proposed. The values obtained by this method showed a strong correlation with true values ( r 2 = 0.89). A possibility of noninvasive quantitative assessment of diastolic filling function was implied.