Empleo de matrices bidimensionales de cámaras de ionización en la verificación de tratamientos de arcoterapia volumétrica modulada

2013 
Volumetric modulated arc therapy (VMAT) requires, as another kind of intensity-modulated radiation therapy (IMRT), patient-specific QA procedures. This work analyzes the method carried out in our institution for VMAT treatment verification. Our hypothesis is that traditional IMRT QA is valid for VMAT technique. Results obtained for absolute point-dose measurements with ion chamber are presented, as well as comparison with treatment planning system calculations (mean difference of (-0.50 ± 0.43)%). In addition, different setups with 2D ion chamber array for dose distributions comparison are analyzed. These detectors are the basis of our QA procedure. Advantages and disadvantages of those setups are shown. The present study includes results for 111 patients treated with VMAT technique from different disease sites. We conclude that 2D ion chamber arrays traditionally used in IMRT QA are valid detectors for rotational techniques if these arrays are used together with additional devices (phantoms, accessories) that allow us to obtain as much information as possible.