A multi-channel acoustics monitor for perioperative respiratory monitoring: preliminary data

2016 
This study pertains to a six-channel acoustic monitoring system for use in patient monitoring during or after surgery. The base hardware consists of a USB data acquisition system, a custom-built six-channel amplification system, and a series of microphones of various designs. The software is based on the MATLAB platform with data acquisition drivers installed. The displayed information includes: time domain signals, frequency domain signals, and tools to aid in the detection of endobronchial intubation. We hypothesize that the above mentioned arrangement may be helpful to the anesthesiologist in recognizing clinical conditions like wheezing, bronchospasm, endobronchial intubation, and apnea. The study also evaluated various types of microphone designs used to transduce breath sounds. The system also features selectable band-pass filtering using MATLAB algorithms as well as a collection of recordings obtained with the system to establish what respiratory acoustic signals look like under various conditions.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    31
    References
    10
    Citations
    NaN