In-office insertion of a miniaturized insertable cardiac monitor: Results from the Reveal LINQ In-Office 2 randomized study

2017 
Background Recent miniaturization of an insertable cardiac monitor (ICM) may make it possible to move device insertion from a hospital to office setting. However, the safety of this strategy is unknown. Objectives The primary objective was to compare the safety of inserting the Reveal LINQ ICM in an office vs a hospital environment. Ancillary objectives included summarizing device- and procedure-related adverse events and responses to a physician questionnaire. Methods Five hundred twenty-one patients indicated for an ICM were randomized (1:1 ratio) to undergo ICM insertion in a hospital or office environment at 26 centers in the United States in the Reveal LINQ In-Office 2 study (ClinicalTrials.gov identifier NCT02395536). Patients were followed for 90 days. Results ICM insertion was successful in all 482 attempted patients (office: 251; hospital: 231). The untoward event rate (composite of unsuccessful insertion and ICM- or insertion-related complications) was 0.8% (2 of 244) in the office and 0.9% (2 of 227) in the hospital (95% confidence interval, −3.0% to 2.9%; 5% noninferiority: P P 15 minutes (16% vs 35%; P P Conclusion The safety profile for the insertion of the Reveal LINQ ICM is excellent irrespective of insertion environment. These results may expand site of service options for LINQ insertion.
    • Correction