TL response of LiF:Mg,Cu,P (GR200A and GR207A) exposed to high-energy 12C ions

2015 
Abstract The characteristics of nat .,7 LiF:Mg,Cu,P thermoluminescent detectors exposed to high-energy 12 C (165 MeV/nucleon) ions and 60 Co γ -rays have been investigated. Through a careful analysis of the glow curves by the GlowFit program, it is found that the relative amplitude of satellite peaks 3, 5, 6 and 7 to peak 4, measured after irradiated with high-energy 12 C ions, is higher than that with 60 Co γ -rays for both types of TL detectors. It is observed that nat .,7 LiF:Mg,Cu,P thermoluminescent detectors would also exhibit a good linear dose response when the delivered dose of high-energy 12 C ions is in the range from 100 mGy to 1600 mGy. Furthermore, the relative efficiency of the nat .,7 LiF:Mg,Cu,P thermoluminescent detectors is independent of the absorbed dose in the linear dose response range. Compared to previously published data, the relative efficiency of the 7 LiF:Mg,Cu,P TL detectors decreases with increasing LET values.