Oxford Optronix MPM 3S : a clinical assessment of a microvascular perfusion monitor

1992 
The Oxford Optronix MPM 3S is a new microvascular perfusion monitor which is promoted as a device for use in the operating theatre. It uses a semiconductor laser diode and applies the Doppler principle to derive a semi-quantitative estimation of microvascular flow. We assessed this instrument with eight healthy volunteers who each performed eight different orlhostalic arm manoeuvres while forearm skin blood flow was monitored. The different manoeuvres caused statistically significant changes in the instrument's reading which generally were consistent with expected changes in blood flow. The monitor also was assessed in the theatre environment with four anaesthetized patients. It proved easy to use, and was not subject to electrical interference from other equipment including short-wave diathermy. The major practical limitation of the technique is the semi-quantitative nature of the measurement. The instrument appears to have potential clinical uses in plastic and vascular surgery.