A calibration method that simplifies and improves accurate determination of peptide molecular masses by MALDI-TOF MS.

2002 
The use of delayed ion extraction in MALDI time-of-flight mass spectrometry distorts the linear relationship between m/z and the square of the ion flight time (t2) with the consequence that, if a mass accuracy of 10 ppm or better is to be obtained, the calibrant signals have to fall close to the analyte signals. If this is not possible, systematic errors arise. To eliminate these, a higher-order calibration function and thus several calibrant signals are required. For internal calibration, however, this approach is limited by signal suppression effects and the increasing chance of the calibrant signals overlapping with analyte signals. If instead the calibrants are prepared separately, this problem is replaced by an other; i.e., the ion flight times are dependent on the sample plate position. For this reason, even if the calibrants are placed close to the sample, the mass accuracy is not improved when a higher-order calibration function is applied. We have studied this phenomenon and found that the relati...