Method design and validation for the determination of uranium levels in human urine using high-resolution alpha spectrometry

2008 
Quantification of uranium in human urine is a valuable technique for assessing occupational and public exposure to uranium. A reliable method has been developed and validated in the ARPANSA Radiochemistry Laboratory by means of standard radiochemical separation and purification techniques and measurement using high-resolution alpha spectrometry. This method can be used to evaluate the levels of naturally occurring 234 U, 235 U and 238 U in urine. Method design and validation is the process of defining an analytical requirement, and then confirming that the method under consideration has performance capabilities consistent with what the application requires. The method was designed to measure levels down to 2 mBq/day of total uranium, corresponding to approximately 1/100th of the annual committed effective dose of 20 mSv. Validation tests were developed to assess selectivity, accuracy, recovery and quantification of uncertainty. The radiochemical recovery of this method was measured using 232 U tracer. The typical minimum detectable concentration for total uranium for 24-h urine samples is approximately 0.6 mBq/day or 0.019 μg/day.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    11
    References
    21
    Citations
    NaN
    KQI
    []