Effects of sodium fluoride mouth rinses on the torsional properties of miniscrew implants

2011 
Introduction Effects of sodium fluoride (NaF) mouth rinse solutions on torsional properties of a miniscrew implant were investigated. Methods As-received Ti-6Al-4V miniscrew implants (AbsoAnchor [Dentos, Inc., Daigu, Korea]) were immersed in 0.1% or 0.2% NaF mouth rinse solution (pH 5.12 and 5.14, respectively) for 1 hour or 24 hours. Miniscrew implants selected as controls were not immersed. Each implant was subjected to increasing torque until fracture (n = 5 in sample groups). Mean moment and twist angle for fracture were compared using 1-way analysis of variance (ANOVA). Surfaces of implants after immersion were observed with a scanning electron microscope (SEM). Electron microprobe and micro-x-ray diffraction analyses were performed to obtain composition information about deposits on implant surfaces. Results Pits and cracks formed on the implant surfaces after immersion in both NaF mouth rinse solutions. Corrosion products, probably sodium aluminum fluoride (Na 3 AlF 6 ), were observed on the implants after immersion in both NaF solutions for both time periods. There were no significant differences for mean torque ( P  = 0.063) and twist angle ( P  = 0.696) at fracture compared with control implants. Conclusions Although titanium alloy miniscrew implants corroded slightly from immersion in 0.1% or 0.2% NaF solutions, mouth rinsing by patients with the same fluoride solution concentrations should not cause deterioration of their torsional performance.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    18
    References
    14
    Citations
    NaN
    KQI
    []