Concentration Monitoring of Plasma Ribavirin: Validation of a Liquid Chromatography-Mass Spectrometric Method and Clinical Sample Collection.

2016 
A liquid chromatography-tandem mass spectrometric (LC-MS/MS) method was developed for routine measurement of ribavirin concentrations in EDTA-anticoagulated plasma.After protein precipitation, we used a bridged ethylene hybrid (hydrophilic interaction) chromatography column, 0.1 mmol/L ammonium formate pH 3.0, and a gradient of 85%-96% acetonitrile to achieve baseline separation of ribavirin from isobaric uridine. Quantitation was assured using both primary (m/z 245.3 > 113.0) and secondary transitions (m/z 245.3 > 96.0) of the protonated species. Chromatographic separation and column washing also negated interference from major phospholipid species.There was a linear relationship between concentration and response to 10 mg/L, with a minimum detectable level and a minimum level of quantitation both of 0.1 mg/L. Imprecision within the assay was 0.4 mg/L. Bias was 6 weeks treatment in 76 patients with hepatitis C.A hydrophilic interaction liquid chromatography-tandem mass spectrometric method to measure ribavirin in plasma was developed. Samples for ribavirin estimation should be kept at 4 °C, separated within 2 hours of collection and stored at 4 °C before analysis, with long-term storage at -20 °C. This method was applied to a study of the ribavirin therapeutic monitoring in patients with hepatitis C.