OBJECTIVE. The purpose of the study was to compare observer performance in the detection of cerebral infarction on a brain CT using medical-grade liquid crystal display (LCD) monitors calibrated with the gray-scale standard display function and with γ 2.2 and using an iPad with a simulated screen setting. MATERIALS AND METHODS. We amassed 97 sample sets, from 47 patients with proven cerebral infarction and 50 healthy control subjects. Nine radiologists independently assessed brain CT on a gray-scale standard display function LCD, a γ 2.2 LCD, and an iPad in random order over 4-week intervals. Receiver operating characteristic (ROC) analysis was performed by using the continuous scale, and the area under the ROC curve (Az) was calculated for each monitor. RESULTS. The Az values for gray-scale standard display function LCD, γ 2.2 LCD, and iPad were 0.875, 0.884, and 0.839, respectively. The difference among the three monitors was very small. There was no significant difference between gray-scale standard di...