A new stem extractor with specific multigrip pliers for revision hip arthroplasty – a biomechanical investigation
2016
IntroductionIt was the aim of this biomechanical study to test a new extractor with 2 specific multigrip pliers on different stem designs and to compare it to a commonly used prefabricated one (UnMod).MethodsThe new extractor comprises 2 specific multigrip pliers, 1 neck-pliers and 1 shoulder-pliers. The tests were performed on a tapered and a nontapered neck stem. They were fixed in specific moulds allowing torque adjustability. The first endpoint was maximum grip with the extractors being tested up to 70 hits or failure (pliers off) in maximally fixed stems. The second endpoint was hits/attempts until stem extraction or failure (pliers off > than 5 times) in moderately fixed stems.ResultsThe best grip on the tapered neck was achieved by the neck-pliers without failure, whereas the others failed consistently (shoulder-pliers: mean 42.9 hits, SD: 3.5/UnMod: 40.1, SD: 5.4; p<0.01). The shoulder-pliers had the best grip on the nontapered neck withstanding more hits (11.9, SD: 1.9) than the others (neck-plie...