Accuracy of GE digital breast tomosynthesis vs supplementary mammographic views for diagnosis of screen-detected soft-tissue breast lesions

2016 
Objective:To compare the accuracy of standard supplementary views and GE digital breast tomosynthesis (DBT) for assessment of soft-tissue mammographic abnormalities.Methods:Women recalled for further assessment of soft-tissue abnormalities were recruited and received standard supplementary views (typically spot compression views) and two-view GE DBT. The added value of DBT in the assessment process was determined by analysing data collected prospectively by radiologists working up the cases. Following anonymization of cases, there was also a retrospective multireader review. The readers first read bilateral standard two-view digital mammography (DM) together with the supplementary mammographic views and gave a combined score for suspicion of malignancy on a five-point scale. The same readers then read bilateral standard two-view DM together with two-view DBT. Pathology data were obtained. Differences were assessed using receiver operating characteristic analysis.Results:The study population was 342 lesion...