Catheter cryoablation of supraventricular tachycardia: results of the multicenter prospective “frosty” trial

2004 
Abstract Objective To study the safety, efficacy, and mapping utility of a new cryoablation catheter. Background The CryoCath Technologies Freezor catheter has been used successfully for cryoablation of supraventricular tachycardia (SVT), but has not been evaluated in a large clinical trial. Methods A multicenter clinical trial to evaluate the safety, efficacy, and cryomapping utility of this cryoablation catheter was conducted in 166 subjects. The target of ablation was the slow pathway in patients with SVT due to AV nodal reentry (AVNRT, n=103), an accessory pathway in patients with AV reentrant SVT (AVRT, n=51) and the AV junction in patients with atrial fibrillation (AF, n=12). Results Acute procedural success (APS) was achieved in 83% of the overall group (95% CI, 76% to 88%). APS in the AVNRT group was 91% (98.3% CI, 82% to 97%), compared to 69% for AVRT (98.3% CI, 51% to 84%) and 67% for AF (98.3% CI, 29% to 93%), a highly significant difference ( P Conclusions Catheter cryoablation of SVT is a safe alternative to RF ablation and is clinically effective in patients with AVNRT. Cryomapping can reversibly identify targets for ablation and can help minimize the risk of inadvertent AV block during ablation.
    • Correction