Upper limit of vulnerability reliably predicts the defibrillation threshold in humans.
1994
BACKGROUNDThe upper limit of vulnerability is the stimulus strength above which electrical stimulation cannot induce ventricular fibrillation even when the stimulus occurs during the vulnerable period of the cardiac cycle. The purpose of this study was to test the hypothesis that the upper limit of vulnerability can accurately predict the defibrillation threshold in patients undergoing implantable cardioverter-defibrillator (ICD) implantation using nonthoracotomy lead systems.METHODS AND RESULTSWe studied 77 patients at the time of ICD implantation. Multiple endocardial-endocardial and endocardial-subcutaneous shock pathways were used. Two different protocols were used to test the upper limit of vulnerability. In protocol 1 (n = 17), the upper limit of vulnerability was tested with two shocks on the peak or the up-slope of the T wave of paced rhythm. The shocks were given randomly either at the peak and 20 milliseconds before the peak of T wave (n = 7) or at 20 and 40 milliseconds before the peak of T wav...