A Comparison of McGrath MAC Versus C-MAC Videolaryngoscopes in Morbidly Obese Patients Undergoing Bariatric Surgery: A Randomized, Controlled Clinical Trial

2019 
Introduction: Airway management in morbidly obese patients is a technical challenge for the anesthesiologists. In this study, we aimed to compare the McGrath MAC with C-MAC videolaryngoscopes for tracheal intubation in morbidly obese patients. Materials and Methods: Eighty morbidly obese patients, scheduled for bariatric surgery, were randomly allocated to two study groups: McGrath-MAC® and C-MAC®. The preoperative airway assessment, incidence and attempts for successful intubation, time to intubation, position for successful intubation, percentage of glottic opening (POGO score), ease of intubation, hemodynamic response, and adverse events of tracheal intubation were recorded. Results: Incidence and attempts for successful intubation and position for successful intubation were similar. Time to intubation was significantly shorter in C-MAC than McGrath (p < 0.001). The POGO scores were similar in both groups (p = 0.057). The heart rate and mean arterial pressure of McGrath was significantly higher than C-...